• Custom torque ranges
  • Symmetrical outputs
  • Easy installation
  • Traceable calibration with certificate included in the standard price
  • Standard 2 year warranty

F315 Low Range Disc Type Static Torque Transducer

Geometry: Slender disc, with outer ring earth fixing and live inner boss usually shaft driven.

This static torque transducer is used extensively in instrumentation and small electric motor test and calibration applications.

Torque transmission must be arranged to input only torque as extraneous forces, such as tare loads or axial loads will cause measurement errors. Note we are able to calibrate with extraneous load cases replicated.

The highly symmetrical signal outputs are usually between 0.2 and 1mV/V for full range.

We are happy to design variants of this transducer to meet your specific requirements. Please consult our engineering department. If you require a higher torque range than 5Nm the F311 is available with ranges up to 2000Nm.

Details of our loadcell families can be found in the Product Guide.

Ordering Codes

See the ordering code sheet for more details. Add range in the required force units.

Most F315 transducers are manufactured to special requirements and are given an F315-Zxxxx number.

F315 Specification

Parameter Value Unit
Non-linearity - Terminal ±0.05 % RT
Hysteresis ±0.1 % RT
Creep - 20 minutes ±0.3 % AT
Repeatability ±0.05 % RT
Rated output - Nominal (depends on range) 0.2 to 1 mV/V
Zero load output ±4 % RT
Temperature effect on rated output per °C ±0.005 % AT
Temperature effect on zero load output per °C (1mV/V) ±0.01 % RT
Temperature range - Compensated -10 to +50 °C
Temperature range - Safe -10 to +80 °C
Excitation voltage - Recommended 10 V
Excitation voltage - Maximum 15 V
Bridge resistance 700 Ω
Insulation resistance - Minimum at 50Vdc 500
Overload - Safe 25 % RT
Overload - Ultimate 50 % RT
Weight - Nominal (excluding cable) 110 to 130 g

The standard ranges are manufactured in aluminium.

Notes

  1. AT = Applied torque.
  2. RT = Rated torque.
  3. Temperature coefficients apply over the compensated range.
  4. Avoid applying axial or side loads to the central boss.

Connections

The loadcell is fitted with 2 metres of PVC insulated 4 core screened cable type 7-2-4C.

Excitation + = Red, Excitation - = Blue, Signal + = Yellow, Signal - = Green, Screen = Orange.

The signal connections are for a positive signal for a clockwise torque on the centre boss. The screen is not connected to the transducer body.

Additional Information

Outline Outline drawing of the product.
STEP file STEP files for standard ranges.
Stock List Loadcell availability.
PDF data-sheet Printable data-sheet. Right click on the link to download the file.

Top